Product Name
Diagnostic Kit for Antibodies to Avian Influenza Virus (ELISA)
Specification
96T×1/kit
Introduction
Avian influenza is also known as real chicken plague or European chicken plague, is caused by the Avian Influenza Virus( AIV ), An acute, highly fatal infectious disease caused by poultry, characterized by acute sepsis death to asymptomatic toxicity.
Principle of Test
The kit is for the qualitative determination of AIV in the sample, adopt AIV antigen to coat microtiter plate, make solid-phase antigen, and then pipette samples to the wells, with anti- goat AIV-ab co
Storage Conditions
u The kit shall be stored at [2-8 ℃].
u The opened Microelisa Stripplate can be stored at [2-8 ℃] and avoid damp. Use for at least 2 months.
Instrument
Mircoplate reader (contain: 450nm, 630nm wavelength), thermostatic equipment (37 degrees Celsius), adjustable Pipette
Composition of the Kit
|
Reagent |
Quantity |
|
|
Microelisa Stripplate |
8well×12strips(1) |
|
|
Positive control |
1×1.0ml |
|
|
Negative control |
1×1.0ml |
|
|
HRP-Co |
1×11ml |
|
|
Sample Diluent |
1×50ml |
|
|
Substrate A solution |
1×6ml |
|
|
Substrate B solution |
1×6ml |
|
|
Stop Solution |
1×6ml |
|
|
Co |
1×40ml |
|
|
User manual |
1 |
|
|
Adhesive Membrane |
1 |
|
|
Sealed bag |
1 |
|
Washing Method
u Manually washing method:
Shake away the remained liquid in the enzyme plates; place some bibulous papers on the test-bed, and flap the plates on the upside down strongly. Inject at least 0.3ml after-dilution washing solution into the well, and marinate 30 second. Repeat this process at 5 times.
u Automatic washing method:
If there is automatic washing machine, it should o
Sample Preparation
u serum:
Coagulation at room temperature for 10-20 min, centrifuge at the speed of 2000-3000 rpm for 20min. Remove supernatant, if precipitation appeared, centrifuge again.
Extract as soon as possible after samples collection, and should be tested as soon as possible after the extraction. If be used recently, kept the samples in 2- 8℃.If not, samples can be kept in -20℃. Avoid repeated freeze-thaw cycles.
u Don’t detect the samples which co
u sample dilution:
40 times dilution(add 5ul sample into 195ul sample diluent, mix)
u 20× Co
Assay Procedure
Step 1: Number: determine the number of well to be used and store unused wells in 4 ℃. Set a blank well without any solution.
Step 2: Prepare sample: pipette Positive co
Step 3: Incubate: Cover with the adhesive Membrane provided, incubate for 30 min at 37℃(recommend water bath) .
Step 4: Co
Step 5: Washing: Uncover the adhesive Membrane, discard liquid, pipette washing buffer to every well, still for 30s then drain, repeat 5 times.
Step 6: Add enzyme: Pipette HRP-Co
Step 7: Incubate: Operation with 3.
Step 8: Washing: Operation with 5.
Step 9: Color: Pipette 50ul Substrate A solution, and then pipette 50ul Substrate B solution to each well, avoid the light preservation for 10 min at 37℃
Step 10: Stop the reaction: Pipette Stop Solution 50μl to each well, stop the reaction (the blue change to yellow).
Step 11: Calculate: take blank well as zero. Read absorbance at 450nm after pipette Stop Solution within 10min.
Calculation of Result
1. Test validity: the average of Positive co
2. A sample value ≥0.38 is positive;
u A sample value between 0.2 and 0.38 for suspicious;
A sample values < 0.2 is negative.
Test Method Limitation
This test is used o
Expiration
12 mo
Attention
² The kit takes out from the refrigeration should be balanced 15-30 minutes in the room temperature, if the coated ELISA plates have not been used up after opening, the plate should be stored in sealed bag.
² Washing buffer will Crystallization separation, it can be heated in water to dissolve.
² Pipette sample with pipettors each step, and proofread its accuracy frequently to avoid the experimental error. Pipette sample within 5 min, if the number of sample is big, recommend using multichannel pipettor.
² The test sample should be kept fresh.
² Adhesive Strip o
² The substrate should evade the light to be preserved.
² Please refer to the user instruction strictly, the test result determination must take the microtiter plate reader as a standard.
² The preparation of samples and all the reagents should refer to infective material process.
² Do not mix reagents with those from other lots.





