当前位置: 给览网 » 图库 » 产品图库
1 / 1

德国Fischer菲希尔RoHS and WEEE ROHS分析仪

日期:2018-01-30    点击:186    查看原图
简介:









 
图库说明
技术参数
 

 

 
FISCHERSCOPE®
X-RAY XAN®
FISCHERSCOPE®
X-RAY XDAL
FISCHERSCOPE®
X-RAY XDV®-SD
测量方向
从下到上
从上到下
从上到下
X射线管为微聚焦管,带铍窗口
 
 
 
高压可调:10kV;30kV;50 kV
 
 
 
开槽测量室
 
 
 
准直器个数:4
 
 
 
Z轴
可编程
可编程
测试台类型
固定工作台
可编程XY工作台
可编程XY工作台
测试点放大倍数
34 -184倍
20 -180倍
20 -180倍
WinFTM® 版本
V6:基本型,带PDM 
WinFTM® Super可选
V6:基本型,带PDM 
WinFTM® Super可选
V6:基本型,带PDM 
WinFTM® Super可选
操作系统:Windows® XP专业版
 
 
 
Amount of primary filters
2
2
6
Digital pulse processor
optional
 
standard

 

 
Descriptions:
Are your products RoHS and WEEE compliant? 

The new EU directives WEEE and RoHS are relevant for all manufacturers of electronic equipment and components the printed circuit industry the cable industry together with all plastics industries. 
From July 2006 the use of certain substances in electrical and electronic instruments will be prohibited.

Which elements does this refer to?

It refers to lead (Pb) mercury (Hg) cadmium (Cd) chromium (Cr VI) polybrominated biphenyls (PBBs) and polybrominated diphenylethers (PBEs). The allowed thresholds are all in the range of 1000 ppm but for cadmium it is just 100 ppm.
Example 1:
  • Handheld Instrument: The measuring result shows: ?RoHS-compliant ?-> WRONG because a mean value is formed over a LARGE area.
  • Fischer Instrument: The measuring result shows: 凬ot Rohs-compliant?-> CORRECT with a spot-precise measurement on the soldering pad.
Example 2:
  • Handheld Instrument: The measuring result shows: ?Not RoHS-compliant ?-> WRONG because the individual components (here the RohS-別xempt?components like diodes) are not considered separately.
  • Fischer Instrument: The measuring result shows: ?RoHS-compliant?-> CORRECT because the spot-precise measurement separates the RoHS-別xempt?diode from the other components.
FISCHERSCOPE?/SUP> X-RAY XAN?/SUP>
  • The reasonable version!
  • Manual positioning via video
  • Very suitable for small parts
FISCHERSCOPE?/SUP> X-RAY XDAL?/SUP>
  • Programmable XY(Z)- table
  • Automatized Scan over areas e.g. individual spots like soldering pads on PCBs or individual components
FISCHERSCOPE?/sup> X-RAY XDV?/sup>-SD
  • Designed especially for RoHS/ WEEE tasks
  • With a digital pulse processor and 6 primary filters
  • With an extremely high repeatability
Further advantages:
  • Full protection instruments with construction approval;
    Thus no endangerment of the operating personnel and no radiation protection agent required.
  • Very small measurement spots possible from about 0.15 mm !
  • Print-out of analyzed area within protocol
  • Only very little experience required
  • Red data warns you when limits are exceeded
  • No time-consuming preparation of specimen
  • Additional advantage: You can also use instruments for coating thickness measurement or solution analysis.
 