联系我们
公司名称:浩瀚色谱(山东)应用技术开发有限公司
联系人:王经理
电话:0632-5667636
手机:15562228838(微信同号)13963221227(微信同号)
传真:0632-5667636
邮件:1404939462@qq.com
地址:山东省滕州市(市场部和技术部);
产品详情

产品简介
测定文具中苯、甲苯、乙苯及二甲苯,芝麻香白酒,3-甲硫基丙醇,室内空气,焦炉煤气,炼厂气,天然气,变压器油,多氯联苯,植物油,增塑剂,塑化剂,氨水,氨气,乙酸乙酯,医疗器械...
产品详细介绍
测定文具中苯、甲苯、乙苯及二甲苯
测定文具中苯、甲苯、乙苯及二甲苯 详细信息:
名称:毛细管柱
规格:60m*0.25mm*1.4um
型号:HH-VRX
应用:文具中苯、甲苯、乙苯及二甲苯的测定方法 气相色谱法
测定文具中苯、甲苯、乙苯及二甲苯 测试谱图:




Determination of benzene, toluene, ethylbenzene and xylene in stationery
Determination of benzene, toluene, ethylbenzene and xylene in statio nery Details:
Name: Capillary Column
Specification: 60m*0.25mm*1.4um
Model: HH-VRX
Application: Determination of benzene, toluene, ethylbenzene and xylene in statio nery gas chromatography
After the sample is prepared as a test sample, it is injected into the gas chromatograph. After the benzene, toluene, ethylbenzene, o-xylene, m-xylene and p-xylene are completely separated from the coexisting substances in the capillary chromatographic column, they are detected by hydrogen flame ionization. Measured by the internal standard method and quantified.
Gas chromatography test conditions:
a) Chromatographic column: polydimethylsiloxane capillary column, 60m×0.25mm inner diameter×1.4μm film thickness;
b) Inlet temperature: 235℃;
c) Column temperature: keep the initial temperature at 50℃ for 1 min, then increase at 8℃/min to 150℃ for 2 min, then increase at 20℃/min to
250℃, keep for 5 min;
d) Detector temperature: 260℃;
e) Carrier gas flow rate: 2.0mL/min
f) Split ratio: split injection, the split ratio is adjustable.
Determination of benzene, toluene, ethylbenzene and xylene in statio nery test spectrum:
如果您觉得“HH-VRX毛细管柱 测定文具中苯、甲苯、乙苯及二甲苯”描述资料不够齐全,请联系我们获取详细资料。(联系时请告诉我从给览网看到的,我们将给您最大优惠!)
本页链接:http://www.geilan.com/com_haohanyiqi/sell/itemid-9127947.html
已经有928位访客查看了本页.
测定文具中苯、甲苯、乙苯及二甲苯 详细信息:
名称:毛细管柱
规格:60m*0.25mm*1.4um
型号:HH-VRX
应用:文具中苯、甲苯、乙苯及二甲苯的测定方法 气相色谱法
样品经制备成测试样后,注入气相色谱仪中,苯、甲苯、乙苯、邻二甲苯、间二甲苯和对二甲苯在毛细管色谱柱中与共存物质完全分离后,用氢火焰离子化检测器测定,以内标法定量。
气相色谱测试条件:
a)色谱柱:聚二甲基硅氧烷毛细管柱,60m×0.25mm 内径×1.4μm 膜厚;
b)进样口温度:235℃;
c)柱温:初始温度 50℃保持 1min,然后以 8℃/min 升 150℃保持 2min,再以 20℃/min 升温
250℃,保持 5 min;
d)检测器温度:260℃;
e)载气流速:2.0mL/min
f)分流比:分流进样,分流比可调。
测定文具中苯、甲苯、乙苯及二甲苯 测试谱图:




Determination of benzene, toluene, ethylbenzene and xylene in stationery
Determination of benzene, toluene, ethylbenzene and xylene in statio
Name: Capillary Column
Specification: 60m*0.25mm*1.4um
Model: HH-VRX
Application: Determination of benzene, toluene, ethylbenzene and xylene in statio
After the sample is prepared as a test sample, it is injected into the gas chromatograph. After the benzene, toluene, ethylbenzene, o-xylene, m-xylene and p-xylene are completely separated from the coexisting substances in the capillary chromatographic column, they are detected by hydrogen flame ionization. Measured by the internal standard method and quantified.
Gas chromatography test conditions:
a) Chromatographic column: polydimethylsiloxane capillary column, 60m×0.25mm inner diameter×1.4μm film thickness;
b) Inlet temperature: 235℃;
c) Column temperature: keep the initial temperature at 50℃ for 1 min, then increase at 8℃/min to 150℃ for 2 min, then increase at 20℃/min to
250℃, keep for 5 min;
d) Detector temperature: 260℃;
e) Carrier gas flow rate: 2.0mL/min
f) Split ratio: split injection, the split ratio is adjustable.
Determination of benzene, toluene, ethylbenzene and xylene in statio
如果您觉得“HH-VRX毛细管柱 测定文具中苯、甲苯、乙苯及二甲苯”描述资料不够齐全,请联系我们获取详细资料。(联系时请告诉我从给览网看到的,我们将给您最大优惠!)
本页链接:http://www.geilan.com/com_haohanyiqi/sell/itemid-9127947.html
已经有928位访客查看了本页.
产品咨询
相关产品推荐
相关技术文章
- icon 甲缩醛纯度及杂质分析专用毛细管柱上市
- icon 热烈祝贺《高纯对二甲苯专用毛细管柱》成功上市
- icon 祝贺《HH-Formol 工业甲醛溶液分析专用毛细管柱
- icon 乙酸中微量甲酸测定专用毛细管柱研制成功
- icon 浩瀚色谱研制的《工业甲醇中微量甲醛测定专用毛








